NickelgridMaterials testing

Method group NG-EIS

Impedance and titration

Impedance separates losses that a capacity curve shows only as one number. Done carelessly it also produces beautiful spectra of a cell that was never at equilibrium, which is why half of this page is about the conditions before the sweep rather than the sweep itself.

NG-EIS-02Spectrum

Potentiostatic impedance

ParameterDefaultNote
Frequency range100 kHz → 10 mHzlower limit on request, 1 mHz costs 2.5 h per spectrum
Points per decade10logarithmic, 7 to 15 accepted
Amplitude10 mV rms5 mV for cells above 2 Ah, linearity checked
States of charge20, 50, 80 %any set between 5 and 100 %
Rest before sweep4 hextended to 12 h if drift exceeds the limit below
Drift limit0.2 mV per hourmeasured over the last 30 minutes of rest
Temperature25 °C−20 to +60 °C, chamber logger quoted

Every spectrum is checked for consistency before it leaves the building. We compute the residual between measured data and its own transform; spectra whose residual exceeds 1 % over the mid-frequency range are repeated rather than fitted. A spectrum that fails twice is reported as failed, with the drift log attached, because a cell that cannot hold a potential for four hours is itself the finding.

Impedance spectrum of an aged graphite cell plotted in the complex plane Two overlapping semicircles between 8 and 38 milliohms followed by a rising tail, typical of film resistance, charge transfer and solid-state diffusion. 0102030405060 0102030 Z′ / mΩ −Z″ / mΩ R₂ = 8.1 mΩ film charge transfer diffusion cell NG-26-118, 25 °C50 % SoC, after 400 cycles
A measured spectrum from method note NG-MN-07. The high-frequency intercept gives the series resistance; the two depressed arcs are not separable by eye and were resolved by fitting, not by reading the plot.
Fitting

Fitting is optional and priced as analysis time, not instrument time. When it is ordered, three things are always stated in the report: the circuit used, the starting values, and the confidence interval on each fitted element. A fit without the last of those is decoration.

  • Default circuit for graphite cells: series resistance, two parallel resistor–capacitance elements with constant-phase behaviour, and a finite-length diffusion element.
  • Elements are only reported when the data actually constrains them. If the lowest measured frequency does not reach the diffusion regime, the diffusion element is left out and we say so.
  • Distribution-of-relaxation-times analysis is available on request. The regularisation parameter changes the number of visible peaks, so we deliver the result at three settings and never at one.
NG-EIS-05Titration

Intermittent titration for diffusion coefficients

A current pulse followed by a long relaxation, repeated across the state-of-charge window. The apparent diffusion coefficient comes out of the voltage transient during each pulse.

Pulse
C/20 for 10 minutes, 25 pulses across the window by default
Relaxation
60 minutes, extended to 120 near phase transitions where relaxation is slow
Sampling
50 ms during the pulse and the first minute of rest, 1 s thereafter
Duration
roughly 30 hours of channel time per cell per direction
Output
coefficient against state of charge, with the particle radius and active mass used in the calculation stated explicitly

The number produced by this method depends on an assumed particle radius and on the assumption of one-dimensional transport into a sphere. For a polycrystalline secondary particle neither holds exactly. We report the coefficient, the inputs, and the factor by which it would change under a different radius assumption — usually within a factor of four. Treat it as a comparison between your own samples, not as a material constant.

Uncertainty
Expanded uncertainty, coverage factor 2, on the impedance bench as calibrated in July 2026.
QuantityUncertaintyDominant contribution
Impedance modulus, 1 Hz to 10 kHz0.4 %reference resistor and lead compensation
Impedance modulus, below 0.1 Hz1.6 %cell drift during the sweep
Phase angle0.3°cabling capacitance
Series resistance0.15 mΩcontact resistance at the holder
Cell temperature0.5 Kgradient between holder and chamber sensor

Figures above 10 kHz are given in the method sheet of the individual job; inductive contributions from the holder and the cell itself dominate there and depend on the format.

Leads are compensated against a short and against a 10 mΩ reference before each session, and the compensation file is delivered with the raw data. The methodology page explains how the uncertainty budget is assembled, and sample requirements covers what we need in the shipment to run this at all.