Cycling (NG-CYC)
NG-CYC-01 - Constant-current cycling at a single temperature
The default cycling request, and the baseline against which the other three methods are defined. One chemistry runs between fixed voltage limits at a single temperature until it reaches a retention floor or a target cycle count. Three cells go on three separate channels so that one misbehaving cell cannot be mistaken for a trend.
- Method code
- NG-CYC-01
- Minimum sample
- 3 coin cells
- Turnaround
- 3–6 weeks
- Temperature
- one setpoint, −20 to +60 °C
- Live section
- the NG-CYC-01 entry in the live catalogue
| Parameter | Value |
|---|---|
| Voltage window | set per chemistry, e.g. 3.0–4.2 V |
| Rate | fixed after formation, quoted in mA |
| Formation | three C/20 cycles before the count starts |
| Reference test | C/20 every 100 cycles |
| Stop criterion | retention floor or cycle count, whichever is first |
- 96-channel cycler, ±5 A per channel
- climate chamber logged independently to ±0.5 K
Current is pinned in milliamperes immediately after formation and is never recalculated as the cell fades, because a falling current gradually treats an ageing cell more kindly and smooths the very curve you are trying to read. Every hundredth cycle is a reference measurement at C/20, which is excluded from the cycle tally but kept in the elapsed-time record. If the spread across the three cells opens past five percent we print all three traces rather than quietly dropping the odd one out.
- per-cycle CSV
- median-and-range summary table
- fade plot
- method sheet
Every method ships with its method sheet, which quotes the code NG-CYC-01 and the exact parameters used for your job.
The NG-CYC-01 sheet is one of nine in the registry on the methods page; cycling fade curves are plotted on the data page.