Impedance (NG-EIS)
NG-EIS-02 - Potentiostatic impedance, 100 kHz to 10 mHz
A spectrum that separates the fast ohmic resistance from the slower charge-transfer arc and from diffusion in the low-frequency tail. The cell is held at a state of charge and nudged by a small ten-millivolt perturbation, kept small enough that the response stays linear. Two cells are measured so that a feature seen in one can be confirmed in the other before anyone believes it.
- Method code
- NG-EIS-02
- Minimum sample
- 2 cells
- Turnaround
- 4 working days
- Temperature
- one setpoint, 25 °C unless stated
- Live section
- the NG-EIS-02 entry in the live catalogue
| Parameter | Value |
|---|---|
| Frequency | 100 kHz down to 10 mHz |
| Amplitude | 10 mV, checked for linearity |
| Points | ten per decade |
| States of charge | as agreed, typically three |
- two impedance stations
- four-wire connection at the cell terminals
We verify the amplitude for linearity before trusting a spectrum, because an over-driven cell returns an arc that looks tidier and means considerably less. Fits are printed with their residuals beside the fitted values, and where an equivalent circuit is used its elements are named rather than buried inside one lumped resistance. The low-frequency tail is the part most often over-interpreted, so we quote the lowest frequency that genuinely reached steady state instead of the nominal ten millihertz.
- Nyquist and Bode spectra
- fitted resistances with residuals
- method sheet
Every method ships with its method sheet, which quotes the code NG-EIS-02 and the exact parameters used for your job.
The NG-EIS-02 sheet is one of nine in the registry on the methods page; cycling fade curves are plotted on the data page.